Relative humidity measurement system

The temperature and humidity detection box of the cold mirror dew point meter was developed by Zhuhai Aobert Co., Ltd., and Oberthur used the British imported high-precision mirror dew point meter to participate in the control of the temperature and humidity test box. The temperature and humidity uniformity and the degree of fluctuation have increased by one level, reaching or even exceeding the technical indicators of the national standard JJG205-2005 temperature and humidity verification box. The perfect combination of mirror dew point meter and temperature and humidity box is realized.


Total accuracy: ±0.15°C dew point; ±0.1°C temperature

Measurement Unit: oC, oF dew point and temperature; %rh; gm-3; gkg-1; aw; t-tdp

Reaction speed: 1 oC sec-1

Power supply: 90~264V, 47/440 Hz. IEC is connected with built-in filter

Transmitter

Working temperature: -40 to +90 oC

* High temperature transmitter option is –20~+130 oC

Mirror: Gold plated copper

Temperature measurement: 1/3 DIN B 4-wire 100W platinum resistance

3mm diameter x 150mm length stainless steel probe

Sample flow rate: 0.1 to 2 N1min-1 (in the sample stop)

Maximum speed: 10 msec-1 (direct plug-in);

30 msec-1 (with sintering protective sleeve)

Pressure: 25 MPa (max)

Cable length: 30 m (max)

Dimensions: 126 length x 48 diameter mm

Weight: 0.5 kg

Display unit

Resolution: The unit is oC, oF and %rh, when it is 0.1, the unit is gm-3 and g/kg-1 is 0.01

Output: Analog: For any parameter, user can select 4~20mA or 0~20mA output

Digital: RS485 (option) or RS232 user selectable baud rate

Analog and digital outputs are updated every 5 seconds

Alarm: 2A when no voltage contact 30VDC;

0.5A at 125VAC; user adjustable

Working environment: Temperature -20~+50 oC; Humidity 98%, no condensation

Outer box: Rectangular boxes with handles 230L x 190W x 85H mm. Panel mounting kits available

Weight: 2.3 kg

Ingress Protection: IP54 (equivalent to NEMA Type 2)

Cable connection: 2m transmitter cable with 10 bayonet connectors

2m temperature probe cable with all-weather garden ring connector

Power cable with IEC connector

Analog/alarm contacts accessible via two terminals

Digital communication is provided by a 9-way D-type connector (installed as a standard part)

RS232 cable 9 way D type plug/socket 2m cable option

High Density Polyethylene Transmitter Sheath; Transmitter Sample Block; Panel Mount Assembly; Transmitter Mounting Nut; Handheld PDA Data Storage Package; Air-Cooled Radiator; Transmitter Additional Cable to 30m;

Unit = mm